|
Vendor Voice
Pipette calibration - Trust your assay results
A pipette is a laboratory instrument used to transport a measured
volume of liquid. It works by creating a vacuum above the liquid-holding chamber
and selectively releasing this vacuum to draw up and dispense liquid.
Pipettes come in several designs for various purposes with
differing levels of accuracy and precision, from single piece glass pipettes
to more complex adjustable or electronic pipettes. Accuracy and precision is
of utmost importance in any pipetting job, however there are many sources of
error affecting pipette calibration which in turn affects the performance of
the pipette.
Simple, reliable pipette calibration maximizes liquid handling
quality. For sustained accuracy and consistent and repeatable operation, pipettes
should be calibrated at periodic intervals.
Accuracy and precision
Inaccuracy is the numerical difference between the average
of a set of measurements and the nominal value. Imprecision means variation
between repeated measurements.
Low imprecision means that calibrator can perform very consistent
calibration (taking into consideration that pipette is in good shape).
COMPONENTS OF QUALITY CALIBRATION
Components of quality calibration can be divided into three
parts: Operator, Equipment and Environment. Impact of these components on pipetting
results are as follows:
Leaky piston / cylinder system (1-50%),
uneven piston movement (0.5%), uneven pipetting rhythm (1.5%), immersion depth
(1.0%), liquid/air temperature variations (0.3%), changes in relative humidity
(3.0%), failure to pre-wet tips (2.0%), failure to "touch off" (3.0%),
poorly sealed pipette tips (0.5-50%), reusing pipette tips (4.0), variations
in air pressure (0.2%), differences in liquid density (1.0%), differences in
vapor pressure (2.0%)
OPERATORS TECHNIQUE
Operator's pipetting technique plays the most important role
in pipettor calibration. The misconception with pipetting is that people consider
it to be a very easy job. But usually this is not the case, especially if the
operator has never been trained for pipetting techniques and calibration. Pipetting
technique must be paid attention to, especially with mechanical pipettors. The
way operators handle plunger movement can make a big difference.
Good pipetting recommendations are as follows:
- Pipettes should be held vertical during aspiration
interval
- Holding a pipette at 30 degrees angle off vertical
position can cause 0.7% more liquid to be aspirated
- Tips should always be pre-wetted
- Immersion depth should be between 2-3mm
- Plunger release should be smooth, methodical and
consistent
- The "Touching off" technique should be
employed when aspirating and dispensing
The graph above shows the difference between experienced
and inexperienced operators.
EQUIPMENT
A) Tips
- To prevent error, tips usage should be as recommended
by the manufacturer
- They should be manufactured from high quality non-wettable
virgin grade polypropylene (clean, hydrophobic surface)
- They should have perfectly centered opening with
smooth interior & straight even sides
- They should be uniform in construction
with a good tip cone match for perfect seal & tip ejection
Also, choose as small a tip as possible, to reduce the air dead space and
ensure that the manufacturing facility is covered by a quality system.
Defects in tips
B) Balance and accessories
- Pipetting error may occur if balance does not have
enough digits. In other words, readability is not sufficient enough (Electronic
balance with resolution 0.001 mg is better recommended, however for calibrating
pipettes above 10ul the balance with resolution 0.01mg is acceptable)
- Balance is old or not calibrated
- Evaporation is not controlled
- Container
Some containers force liquid into a concave configuration
while other force it into a convex configuration.
Glass transfer containers are recommended, since they force
liquid into a concave configuration and help to eliminate the effects of hydrostatic
pressure
- Viscosity of transfer liquid
If liquid other than de-ionized water (ISO 3696) is used,
it will impact calibration results.
ENVIRONMENT
A) Barometric Pressure: Pressure is reduced in accordance
with elevation.
B) Relative Humidity: Accepted guidelines for pipette volume delivery specify
that relative humidity should be between 45%-75%. Humidity traps can be employed
in order to create a "micro-climate" inside the balance, humidity
should be monitored in the room and inside the evaporation trap
C) Temperature: Volume delivery performance is usually defined between 20-25
°C. Stabilisation period of two hours before calibration is recommended
in order to reduce the effects of thermal expansion on pipette, sample and tips.
Pipette, tip and liquid must always be kept at the same temperature.
ISO 8655 - 2 Specification With Maximum Permissible Errors
For Nominal Volumes & Pipette Calibration. Under average conditions GLP
certified labs should check for pipettes calibration every six months to provide
satisfactory performance.
Performance testing
It is important to understand the difference between calibration
and performance testing because most often the performance testing is thought
to be calibration, which it is not. Calibration is the adjustment of the pipette
or the precision instrument to give out the selected volume. Performance testing
is the actual testing procedure, verification of the pipette by using the gravimetric,
photometric or other test methods. In other words, it is the check that the
calibration of the pipette is correct.
Usually, when arriving from the production line, the pipette
fulfils the type specifications given by the manufacturer. The performance test
confirms this. In case the results are incorrect the pipette will need calibration.
If calibration is needed, the pipette will be performance tested again. After
approved results, a quality control certificate with test results is printed
for each individual pipette.
Gravimetric analysis for pipette calibration
In gravimetric calibration you are measuring the weight of
water a pipette dispenses at a given dial setting. This weight must then be
converted into a volume by using the physical property of water density.
Volume (ul ) = Weight of water dispensed ( mg ) / density
of water (mg/ml)
Gravimetric calibration is one of the most accurate ways
to calibrate a pipette and is the accepted standard for the calibration of pipettes.
The technique entails dispensing samples of distilled water into a receiving
vessel perched atop a precision analytical balance. The density of water is
a well-known constant, and thus the mass of the dispensed sample provides an
accurate indication of the volume dispensed. Relative humidity, ambient temperature,
and barometric pressure are factors in the accuracy of the measurement, and
are usually combined in a complex formula and computed as the Z-factor. This
Z-factor is then used to modify the raw mass data output of the balance and
provide an adjusted and more accurate measurement.
There is a clear difference between the performance level
that the pipettor must have when it leaves the factory (controlled testing conditions)
and the performance that is possible to reach by the user in various non-standard
conditions (environmental conditions are not optimum or controlled). In addition
to varying conditions, pipetting techniques may not be adhered to and the pipetted
fluid is seldom purified water.
Conclusions
The trend today is to look for reassurance that the purchased
products fulfill laboratory requirements and meet high end-user expectations.
This often leads to testing or measurement of characteristics against a standard
or specification. To be able to test accurately and reliably, the testing laboratory
should have the full competence to perform calibration according to international
standards. There are two sources of uncertainties in the calibration of pipettors,
namely the uncertainty of the pipettor and the uncertainty of the gravimetric
test method and its related apparatus. An accredited calibration service can
differentiate between these uncertainties and give out the combined uncertainty
of the measurement. The calibration certificate given and signed by the accredited
calibration service separates the uncertainties of the balance and the test
method and, thus, eliminates the sources of error. As a result, the customer
can have full confidence using the pipettors in their tests.
(Article contributed by Transasia Bio-Medicals E-mail: transasia@transasia.co.in)
|